Radiation dose comparison between 70 kVp and 100 kVp with spectral beam shaping for non-contrast-enhanced pediatric chest computed tomography

Meike Weis, Thomas Henzler, John W. Nance, Holger Haubenreisser, Mathias Meyer, Sonja Sudarski, Stefan O. Schoenberg, K. Wolfgang Neff, Claudia Hagelstein

Research output: Contribution to journalArticlepeer-review

55 Scopus citations

Abstract

Objective The aim of this prospective randomized controlled study was to compare 2 techniques for radiation dose reduction in non-contrast-enhanced pediatric chest computed tomography (CT): Low peak kilovoltage imaging at 70 kVp and spectral beam shaping at 100 kVp using a dedicated tin filter (100-kVp Sn). Materials and Methods All chest CT examinations were performed on a third-generation dual-source CT system (SOMATOM Force; Siemens Healthineers, Germany). Fifty children (mean age, 6.8 ± 5.1 years) were examined using the 100-kVp Sn protocol, whereas 25 children received the 70-kVp protocol (mean age, 5.7 ± 5.2 years; 2:1 randomization scheme). Radiation metrics and organ doses were compared between acquisition techniques using commercially available radiation dose analysis software (Radimetrics Inc, Bayer AG, Toronto, Ontario, Canada). Objective image quality, expressed by signal-to-noise ratio and subjective image quality based on a 4-point scale (1, best; 4, worst image quality), were compared. Results Volume CT dose index and size-specific dose estimate were significantly lower in the 100-kVp Sn group compared with the 70-kVp group (0.19 ± 0.12 mGy vs 0.81 ± 0.70 mGy and 0.34 ± 0.13 mGy vs 1.48 ± 1.11 mGy; P < 0.0001 for both). Accordingly, mean effective dose was significantly lower for the 100-kVp Sn examinations (0.21 ± 0.10 mSv) compared with the 70-kVp examinations (0.83 ± 0.49 mSv; P < 0.0001). Calculated organ doses were also significantly lower using the 100-kVp Sn protocol when compared with the 70-kVp protocol; for example, breast dose was reduced by a factor of 4.3. Signal-to-noise ratio was slightly superior for 70-kVp images while lung image quality of the 100-kVp Sn protocol was preferred in subjective analysis (P = 0.0004). Conclusions Pediatric chest CT performed at 100 kVp with an additional tin filter for spectral shaping significantly reduces radiation dose when compared with low peak kilovoltage imaging at 70 kVp and therefore should be preferred in non-contrast-enhanced pediatric chest CT examinations, particularly (given the improved subjective image quality) when the main focus is evaluation of the lung parenchyma.

Original languageEnglish (US)
Pages (from-to)155-162
Number of pages8
JournalInvestigative Radiology
Volume52
Issue number3
DOIs
StatePublished - Mar 1 2017

Keywords

  • chest CT
  • low peak kilovoltage imaging
  • low-dose CT
  • pediatric CT
  • spectral shaping