Bioengineered human acellular vessels for dialysis access in patients with end-stage renal disease: Two phase 2 single-arm trials

Jeffrey H. Lawson, Marc H. Glickman, Marek Ilzecki, Tomasz Jakimowicz, Andrzej Jaroszynski, Eric K. Peden, Alison J. Pilgrim, Heather L. Prichard, Malgorzata Guziewicz, Stanisław Przywara, Jacek Szmidt, Jakub Turek, Wojciech Witkiewicz, Norbert Zapotoczny, Tomasz Zubilewicz, Laura E. Niklason

Research output: Contribution to journalArticlepeer-review

260 Scopus citations

Abstract

Background For patients with end-stage renal disease who are not candidates for fistula, dialysis access grafts are the best option for chronic haemodialysis. However, polytetrafluoroethylene arteriovenous grafts are prone to thrombosis, infection, and intimal hyperplasia at the venous anastomosis. We developed and tested a bioengineered human acellular vessel as a potential solution to these limitations in dialysis access. Methods We did two single-arm phase 2 trials at six centres in the USA and Poland. We enrolled adults with end-stage renal disease. A novel bioengineered human acellular vessel was implanted into the arms of patients for haemodialysis access. Primary endpoints were safety (freedom from immune response or infection, aneurysm, or mechanical failure, and incidence of adverse events), and efficacy as assessed by primary, primary assisted, and secondary patencies at 6 months. All patients were followed up for at least 1 year, or had a censoring event. These trials are registered with ClinicalTrials.gov, NCT01744418 and NCT01840956. Findings Human acellular vessels were implanted into 60 patients. Mean follow-up was 16 months (SD 7·6). One vessel became infected during 82 patient-years of follow-up. The vessels had no dilatation and rarely had post-cannulation bleeding. At 6 months, 63% (95% CI 47-72) of patients had primary patency, 73% (57-81) had primary assisted patency, and 97% (85-98) had secondary patency, with most loss of primary patency because of thrombosis. At 12 months, 28% (17-40) had primary patency, 38% (26-51) had primary assisted patency, and 89% (74-93) had secondary patency. Interpretation Bioengineered human acellular vessels seem to provide safe and functional haemodialysis access, and warrant further study in randomised controlled trials. Funding Humacyte and US National Institutes of Health.

Original languageEnglish (US)
Pages (from-to)2026-2034
Number of pages9
JournalThe Lancet
Volume387
Issue number10032
DOIs
StatePublished - May 14 2016

ASJC Scopus subject areas

  • Medicine(all)

Fingerprint

Dive into the research topics of 'Bioengineered human acellular vessels for dialysis access in patients with end-stage renal disease: Two phase 2 single-arm trials'. Together they form a unique fingerprint.

Cite this