Hip endoprosthesis stabilization with a porous low modulus stem coating: Factors influencing stabilization

William J. Bryan, B. L. McCaskill, H. S. Tullos

Research output: Contribution to journalReview articlepeer-review

15 Scopus citations

Abstract

A polytetrafluoroethylene/vitreous carbon porous material has been investigated as an alternative to acrylic for hip endoprosthesis fixation. This material, which is mechanically fused to the endoprosthesis stem, has been shown to promote soft-tissue ingrowth. In 29 patients receiving a porous coated Thompson endoprosthesis, stabilization of the porous-coated stem as observed by acceptable clinical results occurred in only about 50%. Stabilization of porous coated endoprostheses is unsatisfactory for the followig reasons: lack of press-fit, inadequate prosthetic design, infection, and technical difficulties. The major cause of 40% failure in this series was an initially inadequate press-fit.

Original languageEnglish (US)
Pages (from-to)125-132
Number of pages8
JournalClinical Orthopaedics and Related Research
VolumeNo.157
StatePublished - Jan 1 1981

ASJC Scopus subject areas

  • Surgery
  • Orthopedics and Sports Medicine

Fingerprint

Dive into the research topics of 'Hip endoprosthesis stabilization with a porous low modulus stem coating: Factors influencing stabilization'. Together they form a unique fingerprint.

Cite this