How yarn orientation limits fibrotic tissue ingrowth in a woven polyester heart valve scaffold: A case report

Anne Meddahi-Pelle, Graciela Pavon-Djavid, Nabil Chakfe, Frederic Heim

Research output: Contribution to journalReview articlepeer-review

2 Scopus citations

Abstract

Transcatheter Aortic Valve Implantation (TAVI) has become today a popular alternative technique to surgical valve replacement for critical patients. However, with only six years follow up on average, little is known about the long-term durability of transcatheter implanted biological tissue. Moreover, the high cost of tissue harvesting and chemical treatment procedures favor the development of alternative synthetic valve leaflet materials. In that context, thin, strong and flexible woven fibrous constructions could be considered as interesting candidates. However, the interaction of textile material with living tissue should be comparable to biological tissue, and the Foreign Body Reaction (FBR) in particular should be controlled. Actually, the porosity of textile materials tends to induce exaggerated tissue ingrowth which may prevent the implants from remaining flexible. The purpose of this preliminary animal case study is to investigate the influence of the valve leaflet yarn orientation on the fibrotic tissue ingrowth. For that purpose the in vivo performances of 45° inclined yarn woven valve leaflets implanted in juvenile sheep model were assessed after three months implantation. Results bring out that in the frame of this case study the development of fibrosis is limited with a woven fabric valve obtained from 45° inclined yarns.

Original languageEnglish (US)
Pages (from-to)225-230
Number of pages6
JournalBiomedizinische Technik
Volume66
Issue number2
DOIs
StatePublished - Apr 1 2021

Keywords

  • Foreign body reaction
  • Medical textiles
  • Textile biomaterials
  • Textile heart valve
  • Heart Valve Prosthesis
  • Humans
  • Prosthesis Design
  • Textiles
  • Aortic Valve/physiopathology
  • Polyesters/metabolism
  • Fibrosis/physiopathology
  • Animals
  • Sheep

ASJC Scopus subject areas

  • Biomedical Engineering

Fingerprint

Dive into the research topics of 'How yarn orientation limits fibrotic tissue ingrowth in a woven polyester heart valve scaffold: A case report'. Together they form a unique fingerprint.

Cite this