In vitro and in silico testing of partially and fully bioresorbable vascular scaffold

Nenad Filipovic, Dalibor Nikolic, Velibor Isailovic, Miljan Milosevic, Vladimir Geroski, Georgia Karanasiou, Martin Fawdry, Aiden Flanagan, Dimitrios Fotiadis, Milos Kojic

Research output: Contribution to journalArticlepeer-review

7 Scopus citations

Abstract

Coronary artery disease (CAD), one of the leading causes of death globally, occurs due to the growth of atherosclerotic plaques in the coronary arteries, causing lesions which restrict the flow of blood to the myocardium. Percutaneous transluminal coronary angioplasty (PTCA), including balloon angioplasty and coronary stent deployment is a standard clinical invasive treatment for CAD. Coronary stents are delivered using a balloon catheter inserted across the lesion. The balloon is inflated to a nominal pressure, opening the occluded artery, deploying the stent and improving the flow of blood to the myocardium. All stent manufacturers have to perform standard in vitro mechanical testing under different physiological conditions. In this study, partially and fully bioresorbable vascular scaffolds (BVS) from Boston Scientific Limited have been examined in vitro and in silico for three different test methods: inflation, radial compression and crush resistance. We formulated a material model for poly-L-lactic acid (PLLA) and implemented it into our in-house software tool. A comparison of the different experimental results is presented in the form of graphs showing displacement-force curves, diameter – load curves or diameter - pressure curves. There is a strong correlation between simulation and real experiments with a coefficient of determination (R2) > 0.99 and a correlation coefficient (R) > 0.99. This preliminary study has shown that in-silico tests can mimic the applicable ISO standards for mechanical in vitro stent testing, providing the opportunity to use data generated using in-silico testing to partially or fully replacing the mechanical testing required for regulatory submission.

Original languageEnglish (US)
Article number110158
JournalJournal of Biomechanics
Volume115
DOIs
StatePublished - Jan 22 2021

Keywords

  • Bioresorbable stent
  • Finite element analysis
  • In vitro mechanical test
  • PLLA

ASJC Scopus subject areas

  • Biophysics
  • Orthopedics and Sports Medicine
  • Biomedical Engineering
  • Rehabilitation

Fingerprint

Dive into the research topics of 'In vitro and in silico testing of partially and fully bioresorbable vascular scaffold'. Together they form a unique fingerprint.

Cite this